Cosmic Rays & Scintillation Coincidence Detector

scintillator2

In this post we propose the use of a detector based on a SiPM coupled to a plastic scintillator for the detection of cosmic rays. The detector we are going to use has been described in the previous post SiPM & Plastic Scintillator. To increase the security that logged events are actually caused by cosmic rays we have chosen to use two identical detectors superimposed to each other and connected to a coincidence circuit.

Experimental Setup

In previous posts we have already been described detectors and measurements of cosmic rays :

So we will only describe the new experimental setup we have used.

scintCoinc

As seen in the figure above the two plastic scintillator crystals were placed one above the other and separated by a lead screen to stop environmental radiation.

In the images below you can see the detectors in overlapping position for measurements in coincidence and in the side-by-side position, used to evaluate the random coincidence due to spurious events or background radiation.

scintStackedscintSide

The pulses from SiPM are sent to an electronic device that performs the shaping of the pulses and produces the coincidence pulse with an AND logic gate :

coincCounter

Counting Results

Positioning the scintillator crystals side by side, the result of the counting is practically zero, this is a sign of good temporal “resolution” of the coincidence circuit, in fact the pulses generated by the fast comparator, and used within the coincidence logic, have a duration of only about 200ns; the probability of false coincidences is therefore very low.

By positioning the crystal scintillators one above the other a count value close to the theoretical value is obtained, this value corresponds to the expected value for the cosmic ray flux on the detector surface.

The flow of particles that reach the detector ( at sea level ) should have the following value :

Scintillator crystal area = 32,5cm2
Cosmic particle flux = 0,011 particles / cm²-s-sr
Cosmic particle flux crossing horizontal area = 0,024 particles / cm²-s-sr
32.5 cm2 x 0,024 muons/s cm2 = 0.78 muons/s = 0.78 cps

coincidence

The value obtained is 0,22 CPS which is little lower to the theoretical value of 0,78 CPS.

If you liked this post you can share it on the “social” Facebook, Twitter or LinkedIn with the buttons below. This way you can help us! Thank you !

Donation

If you like this site and if you want to contribute to the development of the activities you can make a donation, thank you !

Check Also

Fluorescence quenching and Stern-Volmer equation

Abstract: in this article we describe some fluorescence quenching measures in some substances. The phenomenon …